Ư¡ (Features)
* The AccuPyc 1330™ Pycnometer
Equipment Description:
The AccuPyc 1330 Pycnometer is a fast, fully automatic density analyzer that performs high-speed, high-precision volume and density measurements on a wide variety of materials including powders, objects, and slurries.
The AccuPyc completes most analyses in 2 to 3 minutes without sacrificing accuracy by using a helium displacement technique for measurement.
Users who require density measurements at a specific temperature can purchasean AccuPyc equipped with a Temperature Control Option.
The AccuPyc 1330TC is designed for connection to a commercial circulating bath and allows for precise temperature control in the 15 to 50 ¡ÆC range.
This unit is equipped with required internal connections, external quick disconnect fittings, and the tubing necessary to connect the unit to an external circulating bath. The temperature inside the sample chamber is stabilized to a fraction of a degree even if the ambient temperature of the laboratory changes.
The AccuPyc 1330TC is available in 10-cc and 100-cc models.
The AccuPyc 1330 GB (Glove Box)
Equipment Description:
This pycnometer, which performs high-speed, high-precision volume and density measurements, is supplied in two sections: the analysis chamber and the electronic control system. The two sections are connected electrically by cables, allowing the analysis chamber to be located in a different environment from the control electronics.
This is useful for those needing to control the instrument from a safe environment while the analysis chamber is located in a radioactive area or in an environment that may be hazardous to the operator. Analysis can be performed on a wide variety of materials including powders, objects, and slurries.
Features
Ability to resolve 0.0001 g/cc
Ultra Precise Calibration
Percent Porosity measurements (requires envelope density)(See GeoPyc 1360 )
Open Cell Volume of foamed materials (See FoamPyc Option)
APPLICATIONS
Foamed materials, Ceramic materials, Metals, Pigments, Plastics, Minerals, Geological cores, Building materials,
Fullerenes, Abrasives, Pharmaceuticals and Slurries
¡¡
* ±Ô°Ý (Specifications)
| Sample Chamber |
Small, 1-cm model: 1.2 cm I.D. x 1.1 cm long (with insert: 0.9 cm I.D x 0.8 cm long)
Standard, 10-cm Model: 1.82cm I.D. x 3.93 cm long
(0.72 in. I.D. x 1.57 in. long)
Large, 100-cm Model: 6.8 cm I.D. x 6.31 cm long (1.92 in I.D. x 2.48 in. long)
|
| Precision: |
Typical reproducibility to within (0.01% of the nominal full-scale sample chamber volume. Guaranteed reproducibility to within )0.02% of the nominal full-scale volume on clean, dry, thermally equilibrated samples. |
| Accuracy: |
Accurate to within 0.03% of reading plus 0.03% of nominal full-scale sample chamber volume. |
| Gases: |
Research grade helium is recommended. If unavailable, use helium with a dewpoint of 67 ¡ÆC -88 ¡ÆF or lower. Carbon dioxide, dry air, nitrogen, etc. can also be used for specific applications. |
| Electrical: |
Voltage: 90 to 264 VAC
Power: 30 VA
Frequency: 50/60 Hz
|
| Physical: |
Width: 53 cm (21 in.)
Height 18 cm (7 in.)
Depth: 36 cm (14 in.)
Weight 19 kg (42 lbs)
|
| Miscellaneous: |
ISO 9001 manufacturer
CE certified
|
|
| CAT No. |
Description |
´ÜÀ§/°¹¼ö |
Price[VATÆ÷ÇÔ] |
|
|
|